Physica

02 Imaging

Scatter-to-primary ratio

SPR = S/P. Grid contrast improvement = (1+SPR)/(1+SPR/K).

Listen

Listen · English

Simulation

Scatter-to-primary ratio — Change the numbers; the scene follows.

Where it works

CT scanner

CT scanner

Gantry

Inside the rotating gantry: tube, bowtie, and detectors that define CTDI and HU.

Open this machine

Formula

SPR=S/P,C=C1+SPR1+SPR/K\mathrm{SPR}=S/P,\quad C'=C\,\frac{1+\mathrm{SPR}}{1+\mathrm{SPR}/K}

Variables

Results

  • SPR

    Scatter-to-primary

    4

  • SPR_out

    SPR after grid

    0.6667

  • C'/C

    Contrast improvement

    3

Explanation

SPR=S/P,C=C1+SPR1+SPR/K\mathrm{SPR}=S/P,\quad C'=C\,\frac{1+\mathrm{SPR}}{1+\mathrm{SPR}/K}

What it means

In a thick abdomen most of the photons reaching the detector are Compton scatter, not primary. SPR of 4–7 is typical without a grid; a well-used grid (selectivity Σ = K, often 5–10) brings it down and restores subject contrast. The price is a Bucky factor of 3–6 in dose. This is why grids are mandatory on Bucky tables and usually retracted on thin extremities and paediatrics. This is a working relation in Diagnostic imaging.

Where it is used

Clinically it sits on the CT scanner — Gantry. Inside the rotating gantry: tube, bowtie, and detectors that define CTDI and HU. Diagnostic equations live on the tube, the detector, and the patient: magnification, air kerma, CTDI, and why bone lights up at 70 kV. They turn a technique chart into physics you can defend.

CT scanner · Open this machine

How to use it

Enter scatter and primary (or SPR directly via S and P) and the grid selectivity K. Read SPR in, SPR out and contrast improvement. K=1 means no grid. Change one input and watch the curve and the simulation follow.

Symbols

  • SScatter4
  • PPrimary1
  • KGrid selectivity6

Worked example

A typical case from the default values: S = 4 (Scatter); P = 1 (Primary); K = 6 (Grid selectivity). Substituting into the relation gives SPR = 4; SPR_out = 0.6667; C'/C = 3. These are teaching numbers — align them with your machine.

Typical values give

  • SPR = 4
  • SPR_out = 0.6667
  • C'/C = 3

Where it comes from

The displayed formula is the working relation. SPR = S/P. Grid contrast improvement = (1+SPR)/(1+SPR/K). Usual reference: Bushberg / AAPM. Derive it in the specialty lesson, then return here to pin the numbers.

Reference: Bushberg / AAPM

Assumptions & limits

Uniform field, no grid cutoff, K independent of energy. Real grids have K that falls at low kV and at wide-angle scatter. Digital scatter correction is a software substitute, not a true K.

Pitfalls

kVp is not the same as effective energy. CTDI is not patient dose — SSDE and organ dose come after. Do not quote DLP as if it were effective dose without a k-factor. Uniform field, no grid cutoff, K independent of energy. Real grids have K that falls at low kV and at wide-angle scatter. Digital scatter correction is a software substitute, not a true K.

Keep this

Technique is physics: kV sets contrast, mAs sets noise, filtration sets the spectrum. Uniform field, no grid cutoff, K independent of energy. Real grids have K that falls at low kV and at wide-angle scatter. Digital scatter correction is a software substitute, not a true K.

In this specialty

Diagnostic imaging