02 Imaging
CT noise scaling
Pixel noise scales as 1/√(mAs · slice · dose).
Listen
Listen · English
Simulation
CT noise scaling — Change the numbers; the scene follows.
Where it works
Radiography room

Bucky / detector
In the Bucky / detector: grid, AEC, DQE, and the pixel that samples the image.
Open this machineFormula
Variables
Results
σ₂
New noise
16.9706HU
σ₂/σ₁
Ratio
1.4142
Curve
Explanation
What it means
In quantum-limited CT, pixel noise σ scales as 1/√(dose × slice thickness × pixel area). Halving mAs multiplies noise by √2 ≈ 1.41. Thinner slices look noisier at the same mAs. This is a working relation in Diagnostic imaging.
Where it is used
Clinically it sits on the Radiography room — Bucky / detector. In the Bucky / detector: grid, AEC, DQE, and the pixel that samples the image. Diagnostic equations live on the tube, the detector, and the patient: magnification, air kerma, CTDI, and why bone lights up at 70 kV. They turn a technique chart into physics you can defend.
Radiography room · Open this machineHow to use it
Use it to predict the noise cost of a low-dose or thin-slice protocol. Iterative reconstruction breaks the simple 1/√N rule — treat the result as an upper bound. Change one input and watch the curve and the simulation follow.
Symbols
- σ₁Original noise12 HU
- mAs₁Original mAs200 mAs
- h₁Original slice5 mm
- mAs₂New mAs100 mAs
- h₂New slice5 mm
Worked example
A typical case from the default values: σ₁ = 12 HU (Original noise); mAs₁ = 200 mAs (Original mAs); h₁ = 5 mm (Original slice); mAs₂ = 100 mAs (New mAs); h₂ = 5 mm (New slice). Substituting into the relation gives σ₂ = 16.9706 HU; σ₂/σ₁ = 1.4142. These are teaching numbers — align them with your machine.
Typical values give
- σ₂ = 16.9706HU
- σ₂/σ₁ = 1.4142
Where it comes from
The displayed formula is the working relation. Pixel noise scales as 1/√(mAs · slice · dose). Usual reference: Bushberg / AAPM. Derive it in the specialty lesson, then return here to pin the numbers.
Reference: Bushberg / AAPM
Assumptions & limits
Quantum noise only. Electronic noise, reconstruction kernel, and AEC along z are omitted. Not valid below the quantum-limited regime (very low dose, small detectors).
Pitfalls
kVp is not the same as effective energy. CTDI is not patient dose — SSDE and organ dose come after. Do not quote DLP as if it were effective dose without a k-factor. Quantum noise only. Electronic noise, reconstruction kernel, and AEC along z are omitted. Not valid below the quantum-limited regime (very low dose, small detectors).
Keep this
Technique is physics: kV sets contrast, mAs sets noise, filtration sets the spectrum. Quantum noise only. Electronic noise, reconstruction kernel, and AEC along z are omitted. Not valid below the quantum-limited regime (very low dose, small detectors).
In this specialty