Optical density
OD = log₁₀(I₀ / I). A factor-of-10 drop in transmission is one OD unit.
Listen
Listen · English
Simulation
Optical density — Change the numbers; the scene follows.
Where it works
Radiography room

Bucky / detector
In the Bucky / detector: grid, AEC, DQE, and the pixel that samples the image.
Open this machineFormula
Variables
Results
OD
Optical density
1.301
I/I₀
Transmission
0.05
Explanation
What it means
Film (and now some display QC) uses logarithmic optical density so that equal perceptual steps are equal OD steps. Base+fog ≈ 0.15–0.20, a chest film mid-grey ≈ 1.2, a black border ≈ 3. H&D characteristic curves plot OD versus log exposure. A 0.3 OD change is a factor of 2 in transmission — coincidentally one doubling of exposure in the linear portion of a film curve (γ ≈ 1). This is a working relation in Diagnostic imaging.
Where it is used
Clinically it sits on the Radiography room — Bucky / detector. In the Bucky / detector: grid, AEC, DQE, and the pixel that samples the image. Diagnostic equations live on the tube, the detector, and the patient: magnification, air kerma, CTDI, and why bone lights up at 70 kV. They turn a technique chart into physics you can defend.
Radiography room · Open this machineHow to use it
Enter incident and transmitted intensity (arbitrary units). OD 1 = 10% transmission, OD 2 = 1%, OD 3 = 0.1%. Change one input and watch the curve and the simulation follow.
Symbols
- I₀Incident intensity100
- ITransmitted intensity5
Worked example
A typical case from the default values: I₀ = 100 (Incident intensity); I = 5 (Transmitted intensity). Substituting into the relation gives OD = 1.301; I/I₀ = 0.05. These are teaching numbers — align them with your machine.
Typical values give
- OD = 1.301
- I/I₀ = 0.05
Where it comes from
The displayed formula is the working relation. OD = log₁₀(I₀ / I). A factor-of-10 drop in transmission is one OD unit. Usual reference: Bushberg / film sensitometry. Derive it in the specialty lesson, then return here to pin the numbers.
Reference: Bushberg / film sensitometry
Assumptions & limits
Incoherent visible light through a film or a printed step wedge. Not a CT number and not a linear grey value of a DR for-processing image (those use a vendor LUT).
Pitfalls
kVp is not the same as effective energy. CTDI is not patient dose — SSDE and organ dose come after. Do not quote DLP as if it were effective dose without a k-factor. Incoherent visible light through a film or a printed step wedge. Not a CT number and not a linear grey value of a DR for-processing image (those use a vendor LUT).
Keep this
Technique is physics: kV sets contrast, mAs sets noise, filtration sets the spectrum. Incoherent visible light through a film or a printed step wedge. Not a CT number and not a linear grey value of a DR for-processing image (those use a vendor LUT).