02 Imaging
Fluoroscopy air-kerma rate
K̇ at IRP from displayed K̇ and inverse-square, plus cumulative KAP → time.
Listen
Listen · English
Simulation
Fluoroscopy air-kerma rate — Change the numbers; the scene follows.
Where it works
Brachytherapy suite

C-arm imaging
At the C-arm image receptor during fluoroscopy — air-kerma rate at the skin.
Open this machineFormula
Variables
Results
K̇_IRP
Rate at IRP
40mGy/min
t
Time for this kerma
12.5min
t_2Gy
Time to 2 Gy
50min
Explanation
What it means
Interventional reference point (IRP) is 15 cm toward the tube from isocentre on a C-arm. Displayed air-kerma rate is legally capped (typical US: 88 mGy/min in normal fluoro, 176 mGy/min in high-level). Skin dose is this number × backscatter × f-factor × a table-attenuation correction, and it is what triggers the 2, 5, 10 Gy substantial-radiation-dose-level alerts. This is a working relation in Diagnostic imaging.
Where it is used
Clinically it sits on the Brachytherapy suite — C-arm imaging. At the C-arm image receptor during fluoroscopy — air-kerma rate at the skin. Diagnostic equations live on the tube, the detector, and the patient: magnification, air kerma, CTDI, and why bone lights up at 70 kV. They turn a technique chart into physics you can defend.
Brachytherapy suite · Open this machineHow to use it
Enter the displayed rate, the distances, and optionally cumulative air kerma. Read IRP rate and the minutes to a 2 Gy skin-dose alert (crude, no BSF). Change one input and watch the curve and the simulation follow.
Symbols
- K̇_refDisplayed rate40 mGy/min
- d_refDisplay distance60 cm
- d_IRPIRP distance60 cm
- KCumulative air kerma500 mGy
Worked example
A typical case from the default values: K̇_ref = 40 mGy/min (Displayed rate); d_ref = 60 cm (Display distance); d_IRP = 60 cm (IRP distance); K = 500 mGy (Cumulative air kerma). Substituting into the relation gives K̇_IRP = 40 mGy/min; t = 12.5 min; t_2Gy = 50 min. These are teaching numbers — align them with your machine.
Typical values give
- K̇_IRP = 40mGy/min
- t = 12.5min
- t_2Gy = 50min
Where it comes from
The displayed formula is the working relation. K̇ at IRP from displayed K̇ and inverse-square, plus cumulative KAP → time. Usual reference: IEC 60601-2-54 / NCRP 168. Derive it in the specialty lesson, then return here to pin the numbers.
Reference: IEC 60601-2-54 / NCRP 168
Assumptions & limits
Does not include backscatter (typically ×1.3 at diagnostic quality), table attenuation, or f-factor to skin (≈ 1.06). A real peak-skin-dose map needs the DICOM RDSR and a geometric model (e.g. em.dose, OpenSkin).
Pitfalls
kVp is not the same as effective energy. CTDI is not patient dose — SSDE and organ dose come after. Do not quote DLP as if it were effective dose without a k-factor. Does not include backscatter (typically ×1.3 at diagnostic quality), table attenuation, or f-factor to skin (≈ 1.06). A real peak-skin-dose map needs the DICOM RDSR and a geometric model (e.g. em.dose, OpenSkin).
Keep this
Technique is physics: kV sets contrast, mAs sets noise, filtration sets the spectrum. Does not include backscatter (typically ×1.3 at diagnostic quality), table attenuation, or f-factor to skin (≈ 1.06). A real peak-skin-dose map needs the DICOM RDSR and a geometric model (e.g. em.dose, OpenSkin).
In this specialty