02 Imaging
Detective quantum efficiency DQE
DQE(f) = SNR_out² / SNR_in² = MTF²(f) / (u² NPS) × (incident quanta).
Listen
Listen · English
Simulation
Detective quantum efficiency DQE — Change the numbers; the scene follows.
Where it works
Radiography room

Bucky / detector
In the Bucky / detector: grid, AEC, DQE, and the pixel that samples the image.
Open this machineFormula
Variables
Results
DQE(0)
Detective quantum efficiency
0.64
DQE
DQE
64%
Explanation
What it means
DQE is the fraction of incoming Poisson information that the detector actually uses. A perfect detector has DQE=1; a real CsI DR panel is 0.6–0.7 at low frequency and falls with MTF². CR is ~0.2–0.3. Raising DQE is how manufacturers cut dose at equal image quality — it is the number to ask for, not just pixel size. This is a working relation in Diagnostic imaging.
Where it is used
Clinically it sits on the Radiography room — Bucky / detector. In the Bucky / detector: grid, AEC, DQE, and the pixel that samples the image. Diagnostic equations live on the tube, the detector, and the patient: magnification, air kerma, CTDI, and why bone lights up at 70 kV. They turn a technique chart into physics you can defend.
Radiography room · Open this machineHow to use it
Enter SNR_out and SNR_in (or incident quanta and measured SNR). DQE at f=0 is this ratio of squares. A panel with SNR_in = 100 (from √N) and SNR_out = 80 has DQE(0) = 0.64. Change one input and watch the curve and the simulation follow.
Symbols
- SNR_outOutput SNR80
- SNR_inInput SNR100
Worked example
A typical case from the default values: SNR_out = 80 (Output SNR); SNR_in = 100 (Input SNR). Substituting into the relation gives DQE(0) = 0.64; DQE = 64 %. These are teaching numbers — align them with your machine.
Typical values give
- DQE(0) = 0.64
- DQE = 64%
Where it comes from
The displayed formula is the working relation. DQE(f) = SNR_out² / SNR_in² = MTF²(f) / (u² NPS) × (incident quanta). Usual reference: IEC 62220 / Bushberg. Derive it in the specialty lesson, then return here to pin the numbers.
Reference: IEC 62220 / Bushberg
Assumptions & limits
Zero-frequency DQE only. Full DQE(f) needs measured MTF and NPS (IEC 62220). Additive electronic noise makes DQE dose-dependent at low exposure.
Pitfalls
kVp is not the same as effective energy. CTDI is not patient dose — SSDE and organ dose come after. Do not quote DLP as if it were effective dose without a k-factor. Zero-frequency DQE only. Full DQE(f) needs measured MTF and NPS (IEC 62220). Additive electronic noise makes DQE dose-dependent at low exposure.
Keep this
Technique is physics: kV sets contrast, mAs sets noise, filtration sets the spectrum. Zero-frequency DQE only. Full DQE(f) needs measured MTF and NPS (IEC 62220). Additive electronic noise makes DQE dose-dependent at low exposure.
In this specialty