01 Therapy
TMR from PDD
Convert percent depth dose to tissue-maximum ratio.
Listen
Listen · English
Simulation
TMR from PDD — Change the numbers; the scene follows.
Where it works
Linear accelerator

Isocenter
At isocenter, on the central axis through the patient (or a phantom in the same place).
Open this machineFormula
Variables
Results
TMR
Tissue-maximum ratio
0.7869
Curve
Explanation
What it means
TMR is the isocentric analogue of PDD: dose at depth in phantom divided by dose at dmax, both at the same SAD. The conversion shown applies an inverse-square factor between the SSD setup and the isocentre. This is a working relation in Radiotherapy.
Where it is used
Clinically it sits on the Linear accelerator — Isocenter. At isocenter, on the central axis through the patient (or a phantom in the same place). Radiotherapy equations sit at the console and in the bunker: output, depth dose, equivalent square, and the monitor units that treat the patient. Hand-calc them beside the TPS, never instead of a commissioned plan.
Linear accelerator · Open this machineHow to use it
Use when you have a PDD table but need TMR for an SAD MU calc. Example: PDD 67%, SSD 100, d=10, dmax=1.5 → TMR ≈ 0.76. Full clinical conversion also uses Sp(surface)/Sp(depth). Change one input and watch the curve and the simulation follow.
Symbols
- PDDPercent depth dose67 %
- SSDSource-to-surface distance100 cm
- dDepth10 cm
- dₘₐₓDepth of dose maximum1.5 cm
Worked example
A typical case from the default values: PDD = 67 % (Percent depth dose); SSD = 100 cm (Source-to-surface distance); d = 10 cm (Depth); dₘₐₓ = 1.5 cm (Depth of dose maximum). Substituting into the relation gives TMR = 0.7869. These are teaching numbers — align them with your machine.
Typical values give
- TMR = 0.7869
Where it comes from
The displayed formula is the working relation. Convert percent depth dose to tissue-maximum ratio. Usual reference: Khan, Ch. 10. Derive it in the specialty lesson, then return here to pin the numbers.
Reference: Khan, Ch. 10
Assumptions & limits
Omits phantom-scatter ratio Sp. Field size at depth differs from the surface; Clarkson or equivalent-square interpolation is still required for blocked fields.
Pitfalls
Never mix PDD from one SSD with TMR from another without converting. Field size at the surface is not the size at isocentre. A hand MU is a check, not a treatment. Omits phantom-scatter ratio Sp. Field size at depth differs from the surface; Clarkson or equivalent-square interpolation is still required for blocked fields.
Keep this
Simplified form without Sp correction. Field size at depth differs from the surface.