01 Therapy
Backscatter factor BSF
BSF = TAR(dmax, s) = dose at dmax in phantom / dose in air at the same point.
Listen
Listen · English
Simulation
Backscatter factor BSF — Change the numbers; the scene follows.
Where it works
Linear accelerator

Isocenter
At isocenter, on the central axis through the patient (or a phantom in the same place).
Open this machineFormula
Variables
Results
BSF
Backscatter factor
1.03
BSF−1
Scatter contribution
0.03
Explanation
What it means
Backscatter is the extra dose at dmax coming from photons scattered back by the phantom. It rises with field size and falls with energy (MV beams have BSF ≈ 1.02–1.06; orthovoltage 5×5 to 20×20 can be 1.1–1.5). Peak scatter factor (PSF) is the same quantity. Output at dmax in phantom = output in air × BSF. This is a working relation in Radiotherapy.
Where it is used
Clinically it sits on the Linear accelerator — Isocenter. At isocenter, on the central axis through the patient (or a phantom in the same place). Radiotherapy equations sit at the console and in the bunker: output, depth dose, equivalent square, and the monitor units that treat the patient. Hand-calc them beside the TPS, never instead of a commissioned plan.
Linear accelerator · Open this machineHow to use it
Enter dose at dmax in phantom and the corresponding in-air dose. For 6 MV 10×10, BSF is only a few percent above 1; for 250 kV it may be 1.3. Change one input and watch the curve and the simulation follow.
Symbols
- D_maxDose at dmax103 cGy
- D_airIn-air dose100 cGy
Worked example
A typical case from the default values: D_max = 103 cGy (Dose at dmax); D_air = 100 cGy (In-air dose). Substituting into the relation gives BSF = 1.03; BSF−1 = 0.03. These are teaching numbers — align them with your machine.
Typical values give
- BSF = 1.03
- BSF−1 = 0.03
Where it comes from
The displayed formula is the working relation. BSF = TAR(dmax, s) = dose at dmax in phantom / dose in air at the same point. Usual reference: Khan / Johns & Cunningham. Derive it in the specialty lesson, then return here to pin the numbers.
Reference: Khan / Johns & Cunningham
Assumptions & limits
Defined at dmax of that energy and field. Electron contamination and tray scatter are not BSF. Do not apply a kV BSF table to an MV beam.
Pitfalls
Never mix PDD from one SSD with TMR from another without converting. Field size at the surface is not the size at isocentre. A hand MU is a check, not a treatment. Defined at dmax of that energy and field. Electron contamination and tray scatter are not BSF. Do not apply a kV BSF table to an MV beam.
Keep this
Name the SSD, energy, and field size with every PDD or TMR you quote. Defined at dmax of that energy and field. Electron contamination and tray scatter are not BSF. Do not apply a kV BSF table to an MV beam.
In this specialty