08 Ultrasound
Ultrasound attenuation
Soft-tissue rule of thumb: 0.5 dB/cm/MHz one way.
Listen
Listen · English
Simulation
Ultrasound attenuation — Change the numbers; the scene follows.
Where it works
Ultrasound

Tissue path
Along the beam in tissue — impedance mismatch, Doppler shift, MI/TI at the focus.
Open this machineFormula
Variables
Results
A
Attenuation
40dB
I/I₀
Relative intensity
0.0001
Curve
Explanation
What it means
Soft-tissue ultrasound attenuation is about 0.5 dB cm⁻¹ MHz⁻¹ one way (1.0 dB round trip). A 5 MHz beam to 8 cm and back loses ~40 dB — TGC exists because of this exponential. This is a working relation in Ultrasound.
Where it is used
Clinically it sits on the Ultrasound — Tissue path. Along the beam in tissue — impedance mismatch, Doppler shift, MI/TI at the focus. Ultrasound equations sit on the probe face and along the beam: impedance, Snell, Doppler, MI and TI. They explain why gel matters, why aliasing appears, and why a mechanical index is on the screen.
Ultrasound · Open this machineHow to use it
Set round-trip = 1 for pulse-echo. Fat ~0.6, muscle ~1.0, bone much higher. Frequency is a trade: penetration vs resolution. Change one input and watch the curve and the simulation follow.
Symbols
- αAttenuation coefficient0.5 dB/cm/MHz
- fFrequency5 MHz
- zDepth (one way)8 cm
- roundRound trip? 1 = yes1
Worked example
A typical case from the default values: α = 0.5 dB/cm/MHz (Attenuation coefficient); f = 5 MHz (Frequency); z = 8 cm (Depth (one way)); round = 1 (Round trip? 1 = yes). Substituting into the relation gives A = 40 dB; I/I₀ = 0.0001. These are teaching numbers — align them with your machine.
Typical values give
- A = 40dB
- I/I₀ = 0.0001
Where it comes from
The displayed formula is the working relation. Soft-tissue rule of thumb: 0.5 dB/cm/MHz one way. Usual reference: Bushberg. Derive it in the specialty lesson, then return here to pin the numbers.
Reference: Bushberg
Assumptions & limits
Constant α, no focusing gain, no harmonic generation. Real TGC also compensates beam spreading.
Pitfalls
Soft-tissue 1540 m/s is an assumption — not a measurement in that patient. Doppler angle 90° gives no shift. MI and TI are on-screen estimates, not absorbed dose. Constant α, no focusing gain, no harmonic generation. Real TGC also compensates beam spreading.
Keep this
Use 2z for round trip. α ≈ 0.5 dB cm⁻¹ MHz⁻¹.
In this specialty