Physica

06 Dose

Temperature-pressure correction

Open-air chamber correction to 22 °C and 101.33 kPa.

Listen

Listen · English

Simulation

Temperature-pressure correction — Change the numbers; the scene follows.

Where it works

Water phantom

Water phantom

Ion chamber

In the water tank under the linac, at the ion chamber — reference dosimetry happens here.

Open this machine

Formula

kTP=273.2+T295.2101.33Pk_{TP}=\frac{273.2+T}{295.2}\cdot\frac{101.33}{P}

Variables

Results

  • k_TP

    Temp/pressure factor

    1

Explanation

kTP=273.2+T295.2101.33Pk_{TP}=\frac{273.2+T}{295.2}\cdot\frac{101.33}{P}

What it means

Open-air ionisation chambers are calibrated at 22 °C and 101.33 kPa (TG-51). Density of the cavity gas scales as P/T, so k_TP = (273.2+T)/295.2 × 101.33/P. A hot, high-altitude clinic has k_TP > 1 (less gas, you multiply the reading up). This is a working relation in Dosimetry.

Where it is used

Clinically it sits on the Water phantom — Ion chamber. In the water tank under the linac, at the ion chamber — reference dosimetry happens here. Dosimetry is the chamber in water under the linac, or the well counter in the hot lab: TG-51, TRS-398, kerma, and recombination. These numbers are the calibration the rest of the department borrows.

Water phantom · Open this machine

How to use it

Measure T in the phantom (not the room) after equilibrium. P is station pressure, not corrected to sea level. A 5 °C error is 1.7% in dose — not negligible. Change one input and watch the curve and the simulation follow.

Symbols

  • TTemperature22 °C
  • PPressure101.33 kPa

Worked example

A typical case from the default values: T = 22 °C (Temperature); P = 101.33 kPa (Pressure). Substituting into the relation gives k_TP = 1. These are teaching numbers — align them with your machine.

Typical values give

  • k_TP = 1

Where it comes from

The displayed formula is the working relation. Open-air chamber correction to 22 °C and 101.33 kPa. Usual reference: AAPM TG-51. Derive it in the specialty lesson, then return here to pin the numbers.

Reference: AAPM TG-51

Assumptions & limits

Ideal-gas, dry air. Humidity correction is usually <0.3% and often omitted. Sealed chambers (some Farmer-type with a collector seal) should not receive k_TP blindly.

Pitfalls

kQ is for that chamber and that beam quality — not a neighbour's value. Polarity and recombination are measured, not copied. A ⁶⁰Co N_D,w is not an MV calibration until kQ is applied. Ideal-gas, dry air. Humidity correction is usually <0.3% and often omitted. Sealed chambers (some Farmer-type with a collector seal) should not receive k_TP blindly.

Keep this

Trace every gray back to a protocol, a chamber, and a quality index. Ideal-gas, dry air. Humidity correction is usually <0.3% and often omitted. Sealed chambers (some Farmer-type with a collector seal) should not receive k_TP blindly.

In this specialty

Dosimetry