Physica

06 Dose

Spencer–Attix cavity dose

D_med = D_gas × (L̄/ρ)_gas^med × P. Restricted stopping-power ratio with cutoff Δ.

Listen

Listen · English

Simulation

Spencer–Attix cavity dose — Change the numbers; the scene follows.

Where it works

Water phantom

Water phantom

Ion chamber

In the water tank under the linac, at the ion chamber — reference dosimetry happens here.

Open this machine

Formula

Dmed=Dgas(Lˉρ)gasmedPflPwallD_{\mathrm{med}}=D_{\mathrm{gas}}\left(\frac{\bar L}{\rho}\right)_{\mathrm{gas}}^{\mathrm{med}}P_{fl}P_{\mathrm{wall}}

Variables

Results

  • D_med

    Medium dose

    0.01119Gy

Explanation

Dmed=Dgas(Lˉρ)gasmedPflPwallD_{\mathrm{med}}=D_{\mathrm{gas}}\left(\frac{\bar L}{\rho}\right)_{\mathrm{gas}}^{\mathrm{med}}P_{fl}P_{\mathrm{wall}}

What it means

Bragg–Gray assumes the cavity does not perturb the electron fluence and uses unrestricted stopping powers. Spencer–Attix is the practical version: δ-rays above a cutoff Δ (matched to cavity size, ~10 keV for a Farmer) are treated as part of the electron field, and restricted stopping powers L_Δ are used. Every TG-51 / TRS-398 kQ is, under the hood, a Spencer–Attix ratio evaluated in Monte Carlo. This is a working relation in Dosimetry.

Where it is used

Clinically it sits on the Water phantom — Ion chamber. In the water tank under the linac, at the ion chamber — reference dosimetry happens here. Dosimetry is the chamber in water under the linac, or the well counter in the hot lab: TG-51, TRS-398, kerma, and recombination. These numbers are the calibration the rest of the department borrows.

Water phantom · Open this machine

How to use it

Enter dose in gas (from J / m, with W/e), the restricted stopping-power ratio, and a combined perturbation P (Pwall Pfl Pgr Pcel…). Read D_med. A Co-60 Farmer in water has (L̄/ρ) ≈ 1.127 and P ≈ 0.99. Change one input and watch the curve and the simulation follow.

Symbols

  • D_gasGas dose0.01 Gy
  • (L̄/ρ)Stopping-power ratio1.127
  • PPerturbation product0.993

Worked example

A typical case from the default values: D_gas = 0.01 Gy (Gas dose); (L̄/ρ) = 1.127 (Stopping-power ratio); P = 0.993 (Perturbation product). Substituting into the relation gives D_med = 0.01119 Gy. These are teaching numbers — align them with your machine.

Typical values give

  • D_med = 0.01119Gy

Where it comes from

The displayed formula is the working relation. D_med = D_gas × (L̄/ρ)_gas^med × P. Restricted stopping-power ratio with cutoff Δ. Usual reference: Spencer & Attix 1955 / Attix. Derive it in the specialty lesson, then return here to pin the numbers.

Reference: Spencer & Attix 1955 / Attix

Assumptions & limits

Δ is not an input — you must supply an (L̄/ρ) that already corresponds to a chosen Δ. Fano’s theorem says a density (not Z) perturbation of a homogeneous medium does not change fluence; real chambers violate that via the wall and electrode.

Pitfalls

kQ is for that chamber and that beam quality — not a neighbour's value. Polarity and recombination are measured, not copied. A ⁶⁰Co N_D,w is not an MV calibration until kQ is applied. Δ is not an input — you must supply an (L̄/ρ) that already corresponds to a chosen Δ. Fano’s theorem says a density (not Z) perturbation of a homogeneous medium does not change fluence; real chambers violate that via the wall and electrode.

Keep this

Trace every gray back to a protocol, a chamber, and a quality index. Δ is not an input — you must supply an (L̄/ρ) that already corresponds to a chosen Δ. Fano’s theorem says a density (not Z) perturbation of a homogeneous medium does not change fluence; real chambers violate that via the wall and electrode.

In this specialty

Dosimetry