Physica
Lessons

01 Therapy

From isocenter to monitor units

Inverse square, percent depth dose, and the MU you will actually type.

For Planning physicists · RTTs·9 min

Listen

0:000:00

Loading studio reading…·Listen

You will be able to

  1. 01Move a point source and see dose rate change as 1/r².
  2. 02Read a PDD: dmax, the fall, energy and field size.
  3. 03Write an SSD MU as dose over output times PDD times modifiers.

Chapter 01

The point that does not move

Isocenter is where gantry, collimator and couch axes meet. Inverse square is the geometry of that point. Double the distance, quarter the intensity — in air, from a point. Inside a patient you still start here, then you add attenuation and scatter.

SSD changes are inverse-square first, scatter second.

Simulation

Inverse square law — Change the numbers; the scene follows.

Variables

I₂ = 156.25 cGy/min

I2=I1(d1d2)2I_2 = I_1\left(\frac{d_1}{d_2}\right)^2

Keep thisSSD changes are inverse-square first, scatter second.

Open the calculator — Inverse square law

Chapter 02

Walking the axis

Percent depth dose is central-axis dose at depth, as a percent of dose at dmax, at a stated SSD. Electrons contaminate the surface; photons build to dmax; then the curve falls. Higher energy pushes dmax deeper. Larger fields lift the tail with scatter. Never mix PDDs from two SSDs without Mayneord or a TMR conversion.

A PDD is married to its SSD, energy, and field.

Simulation

Percent depth dose — Change the numbers; the scene follows.

Variables

PDD = 90 %

PDD(d)=D(d)D(dmax)×100\mathrm{PDD}(d)=\frac{D(d)}{D(d_{\max})}\times 100

Keep thisA PDD is married to its SSD, energy, and field.

Open the calculator — Percent depth dose

Chapter 03

What you type

Monitor units for an SSD setup are prescribed dose divided by calibration output, PDD over 100, and the modifiers Sc, Sp, wedge, tray. If the machine is one centigray per MU at dmax, ten by ten, reference SSD, this formula is the daily workhorse. Always compare with the record-and-verify.

Simulation

Monitor units — SSD technique — Change the numbers; the scene follows.

Variables

MU = 289.76 MU

Keep thisAn MU is not a dose. It is a charge the monitor chamber counted.

Open the calculator — Monitor units — SSD technique

In this specialty

Open this machineLinear accelerator